Microplaatlezer, Varioskan™ LUX

Leverancier: THERMO LABSYSTEMS LIFE SCIENCE
VLBLATD1 VLBLATD0 VLBLATD2 VLBLA0D2 VLBLA0D1 VLBLA0D0 VLBL0TD2 VLBL0TD1 VLBL0TD0 VLBL00D2 VLBL00D1 VLBL00D0 VLB000D2 VLB000D1 VLB000D0 VL0LATD2 VL0LATD1 VL0LATD0 VL0LA0D2 VL0LA0D1 VL0LA0D0 VL0L0TD2 VL0L0TD1 VL0L0TD0 VL0L00D2 VL0L00D1 VL0000D2 VL0000D1 VL0000D0 VL0L00D0 VLBL0TGD0 VLB000GD0 VLB000GD1 VLB000GD2 VLBL00GD0 VLBL00GD1 VLBL00GD2 VLBL0TGD1 VLBL0TGD2 VLBLA0GD0 VLBLA0GD1 VLBLA0GD2 VLBLATGD0 VLBLATGD1 VLBLATGD2
735-0323EA 63200 EUR
735-0323 735-0322 735-0324 735-0321 735-0320 735-0319 735-0318 735-0317 735-0316 735-0315 735-0314 735-0313 735-0312 735-0311 735-0310 735-0309 735-0308 735-0307 735-0306 735-0305 735-0304 735-0303 735-0302 735-0301 735-0300 735-0299 735-0298 735-0297 735-0296 735-0295 LABLVLBL0TGD0 LABLVLB000GD0 LABLVLB000GD1 LABLVLB000GD2 LABLVLBL00GD0 LABLVLBL00GD1 LABLVLBL00GD2 LABLVLBL0TGD1 LABLVLBL0TGD2 LABLVLBLA0GD0 LABLVLBLA0GD1 LABLVLBLA0GD2 LABLVLBLATGD0 LABLVLBLATGD1 LABLVLBLATGD2
Microplaatlezer, Varioskan™ LUX
Microplaatlezers
Varioskan™ LUX comes equipped with a range of measurement technologies including absorbance, fluorescence intensity and FRET as standard, and with optional luminescence, AlphaScreen and time-resolved fluorescence (TRF) modules. The instrument selects the measurement wavelength using filters or monochromators, depending on which is optimal for each measurement technology.

  • Automatisch dynamisch bereik selecteert het optimale leesbereik op basis van de signaalintensiteit
  • Slimme veiligheidscontroles helpen het instrument en monsters te beschermen tegen gebruikersfouten
  • Autokalibratie en zelfdiagnose voor vertrouwen in resultaten
  • Nauwkeurige temperatuurregeling voor temperatuurkritische toepassingen
  • Geïntegreerde gasregelmodule (optioneel) voor het instellen van CO₂- en O₂-concentraties voor celgebaseerde assays
  • Optionele dispensers voor gelijktijdige afgifte en meting voor signaalbewaking aan het begin van een reactie

SkanIt-software (beschikbaar in 9 talen):
Intuïtieve gebruikersinterface voor een vereenvoudigde meetopstelling
Virtual pipette™-hulpmiddel voor eenvoudig definiëren van monsters op basis van de opmaak van de plaat
Ingebouwde berekeningsopties voor een eenvoudigere gegevensverwerking
Handmatig of automatisch gegevens exporteren naar elke locatie
Meetgegevens worden continu opgeslagen in de database; hiermee wordt gegevensverlies als gevolg van onderbrekingen, zoals bij een stroomstoring of per ongeluk afsluiten, voorkomen
Verkrijgbaar in 2 edities: de Research Edition (meegeleverd) of Drug Discovery Edition

Absorbance:
Plate types: 6 to 384 well plates
Wavelength selection: Double monochromators
Wavelength range: 200 to 1000 nm
Light source: Xenon flash lamp
Read out range: 0 to 6 Abs
Linear measurement range: 0 to 4 Abs (96-well plate) at 450 nm, ±2%; 0 to 3 Abs (384-well plate) at 450 nm, ±2%
Accuracy: 0,003 Abs or ±2%, at 200 to399 nm (0 to 2 Abs); 0,003 Abs or ±1%, at 400 to 1000 nm (0 to 3 Abs);
Precision: SD <0,001 Abs or CV <0,5%, at 450 nm (0 to 3 Abs)

Fluorescence intensity:
Plate types: 6 to 1536 well plates
Wavelength selection: Double excitation and emission monochromators
Excitation wavelength range: 200 to 1000 nm
Emission wavelength range: 270 to 840 nm
Light source: Xenon flash lamp
Sensitivity: Top reading: <0,4 fmol fluorescein/well (black 384-well plate); Bottom reading: <4 fmol fluorescein/well, (clear bottom black 384-square well plate)
Dynamic range: Top reading >6 decades; Bottom reading >5,5 decades

Time-resolved fluorescence:
Plate types: 6 to 1536 well plates
Wavelength selection: Filters (spectral scanning with double excitation and emission monochromators)
Excitation wavelength range: Fixed to 334 nm (spectral scanning 200 to 840 nm)
Emission wavelength range: 400 to 700 nm (spectral scanning 270 to 840 nm)
Light source: Xenon flash lamp
Sensitivity: <1 amol Eu/well (white low volume 384 well plate)
Dynamic range: >6 decades

Luminescence:
Plate types: 6 to 1536 well plates (spectral scanning 6 to 384 well plates)
Wavelength selection: Direct or filters (spectral scanning with double monochromators)
Wavelength range: 360 to 670 nm
Sensitivity: <7 amol ATP/well (white 384-well plate)
Dynamic range: >7 decades

AlphaScreen:
Plate types: 6 to 1536 well plates
Wavelength selection: Filters
Excitation wavelength range: Fixed to 680 nm
Emission wavelength range: 400 to 660 nm
Light source: LED
Sensitivity: <100 amol phosphotyrosine/well (white 384-well plate)

Dispensing:
Plate types: 6 to 384 well plates
No. of dispensers: None, one or two
Syringe size: 1 ml (standard), 5 ml (optional)
Dispensing volume: 2 to 5000 μl, in 1 µl increments (1 ml syringe); 5 to 25 000 μl, in 5 µl increments (5 ml syringe)
Accuracy: <1 μl with 50 μl (0,4 mm tip), <0,2 μl with 5 μl (0,25 mm tip)
Precision: <1 μl with 50 μl (0,4 mm tip), <0,25 μl with 5 μl (0,25 mm tip)
Dead volume: Reagent loss <100 μl; Total tubing volume <800 μl

Bestelinformatie: Neem contact op met Avantor voor meer technische specificaties met betrekking tot alle beschikbare modules voor de Varioskan® LUX en ingebouwde dispensers, incubator en schudder.
Bestel nu

Learn more

About VWR

Avantor is a vertically integrated, global supplier of discovery-to-delivery solutions for...

Meer About VWR